Electrical Specifications | ||
Voltages Available | 115 Volts 60 Hertz 1 Phase | |
Amperage | 15 Amp Dedicated Line | |
12 Running Amps | ||
NEMA Configuration | NEMA 5-15 | |
Certification(s) | ETL Listed | |
Control Specifications | ||
Temperature Control | Digital Display | |
Min / Max Temperature | Display and Reset | |
Temperature Alarm | Audible High / Low Alarm | |
Process Phase Specific | ||
Additional Alarm Types | Power Failure, Sensor Failure | |
Programmable | Ramp and Soak | |
Event Logger | Stores 100 Previous Events | |
Clock | Real Time Clock | |
Data Connections | RS-485 (MODBUS-RTU) | |
Data Download | Via, USB | |
Battery Backup | Only Powers Alarm | |
Alarm Relay | Dry Contacts | |
Temperature Adjustment | ± 0.1°C | |
Temperature Stability | ± 0.5°C From 4°C to 70°C (Range Dependent) | |
Temperature Uniformity | ±1.0 °C From 4°C to 70°C (Range Dependent) | |
General Specifications | ||
Air Circulation | Ducted Airflow | |
Plenum System | ||
Compressor(s) | Hermetic Compressor | |
Refrigeration | R600a | |
Insulation | Polyurethane | |
Opening | 1 Front Door | |
Opening Type | Hinged | |
Door Material | Insulated Steel | |
Lockable Door | Keyed Lock | |
Access Porthole | 3/4" Diameter | |
Storage | 3 Adjustable Stainless Tray Shelves | |
Exterior | Powder Coated Steel | |
Interior | Stainless Steel | |
Interior Lighting | Adjustable LED | |
Casters | 4, Locking | |
Warranty | 18 Months Parts and Labor | |
5 years on Compressor Parts (No Labor) |
Manufacturer | So-Low |
Capacity [ Cubic Feet ] | 23 Cubic Feet |
Capacity [ Liters ] | 651 Liters |
Temperature Range [ °C ] | 4°C to 70°C |
Temperature Range [ °F ] | 39°F to 158°F |
Dimensions | [ W x D x H ] |
Exterior [ Inches ] | 27" x 36.5" x 81.5" |
Interior [ Inches ] | 22" x 25.5" x 59" |
Exterior [ Centimeters ] | 68 x 92 x 207 |